Using printing solutions
A printing session requires two solutions:
- chamber solution
- iontip ink
Each solution should be handled accordingly to local regulations wearing the proper safety equipment.
Equipment needed
- 0.1-5 ml, 1-100 µl and 0.1-10 µl micropipettes
- 2 ml syringe
- 200 nm (or less) pore size filters (hydrophilic)
- 2 ml vials
Printing Copper
Using the chamber solution [Cu]
The chamber solution [Cu]
can be used as received. 10 to 12 ml of the solution can be pipetted into the assembled printing chamber or replaced when needed. The chamber solution [Cu]
should be changed daily. The used solution must be disposed according to local regulations. Remember to also setup the PSTrace software if not done already, applying a potential of -0.5 V.
Using the iontip ink [Cu]
The iontip ink [Cu]
must be filtered before use using the provided syringes and filters.
When using the iontip ink brightener [Cu]
, 20 µl of each component (A, B and C) must be added to 1 ml of iontip ink [Cu]
using the provided micropipette. Micropipette tips are single use and must be changed for each component. The mixed iontip ink brightener [Cu]
filtered before use using the provided syringes and filters.
The probe can then be filled with 0.2 to 1 µl of filtered ink.
Solution supplies
Solutions described here can be ordered at Exaddon or via your local distributor. A step by step guide on how to prepare basic versions of the chamber solution [Cu]
and iontip ink [Cu]
can be found here.