Accessories
Iontips
Used for XY calibration, printing or simple as protection for transporting or storing the printing head.
Bottom view of consumable including cantilever
Side view of consumable incl. tip geometry
Consumable holder
Used for calibration and iontips gripping and dropping.
Warning
Use only A1, A3, C1 and C3 slots! Always store probes back in their blister or in the holder rack mounted on a wellplate before removing the consumable holder from the printer.
Consumable holder rack for storage
Placed on top of a wellplate filled with a storage liquid, tips can be stored in each slot for hours. The ink reservoir tends to dry first: to avoid it, the holder rack can be covered with a lint free wipe kept humid, closed with the wellplate lid.
Wellplate
Used to hold various liquids (water, electrolyte, etchant,…) for setup and cleaning purposes. The wellplate used must suit the wellplate defined in the software. To use other wellplates, their new plate definition must be defined.
Calibration plate
Used during the installation, calibration and maintenance step. Please store it properly to avoid dust deposition on the plate.
Printing chamber
Printing chamber holds the printing cell for electrochemical printing processes. To clean the carbon parts, Teflon parts and O-rings in contact with the electrolyte, washing with deionized water is usually sufficient if the same supporting electrolyte is used. For in-depth washing, please refer the chamber cleaning procedures.
To remove the substrate, turn by two steps the sample with the tool:
Spare sealing screws and O-rings
Spare parts are delivered with the system:
- noze O-rings (head side)
- consumable O-rings (probe side)
- a spare sealing screw already assembled with a consumable O-ring.
Probe lever
A small steel needle used to open the grippers of the printing head and to release or grip consumables by hand during installation or maintenance steps.